Analysis preferences and defaults

Purpose

Open File > Settings Center to review Program settings, Global Analysis defaults, and analysis profiles. A profile records software choices; it does not establish that a choice is scientifically correct for every muscle, species, amplifier, sampling rate, or protocol. Validate and record the profile used for each analysis.

Core defaults

Setting Shipped default What it changes
Stimulus bin size 0.01 V Groups nearby stimulus voltages for aggregate curves. Changing it can change means, error bars, and M-max inputs.
Displayed post-stimulus time 1000.0 ms Default time shown after the stimulus in plots; shorter recordings remain limited to their acquired width.
Displayed pre-stimulus time 250.0 ms Default baseline shown before the stimulus; shorter recordings remain limited to their acquired width.
Default amplitude method RMS The starting method for plots; choose the method required by the analysis plan.
Default channel names LG, TA, SOL Labels used when the imported data do not provide channel names.

EMG filter defaults

Setting Shipped default Review before use
Low cutoff 100 Hz It should retain the signal content you need.
High cutoff 3500 Hz It must remain below the Nyquist frequency for the recording.
Filter order 4 Changing it alters the transition region and transient behavior.

MonStim uses forward-and-reverse filtering to avoid net phase shift. It does not silently substitute invalid cutoffs. See EMG processing and transformations for the signal path.

M-max defaults

Setting Shipped default What to know
Largest plateau run 15 points The first plateau length MonStim tests.
Smallest plateau run 2 points The shortest run it will consider after no larger run qualifies. Two points are weak evidence.
Plateau variation threshold 0.15 How much the M-wave values may vary within a plateau, relative to the high-response signal level. 0.15 allows up to 15% variation.
Validation tolerance 1.05 Allows a candidate up to 5% above the relevant mean.
Smoothing-window ratio 0.25 Sets the Savitzky–Golay window from the number of stimulus levels.
M-wave recognition names M-wave aliases Global names that identify the M-response latency window for M-max. The user list replaces the shipped aliases; an empty list disables automatic recognition.

The M-wave recognition names are global user settings, not profile settings. Edit them in File > Settings Center > Global Analysis > Latency windows. The M-max estimator also uses a third-order smoothing polynomial, a 95th-percentile candidate, and a top-20% candidate. It reports M-max as unavailable when no plateau is detected. These are reproducibility settings, not physiological constants. See M-max estimation before using M-max normalization.

Presets and appearance

Latency-window presets are templates for their named protocols, not universal timing recommendations. Colors and plot-style settings change presentation, not calculated amplitudes. A Dataset- or Experiment-level latency-window application copies the selected draft to included child sessions; review the target carefully before applying.